Observational study measures aerosol characteristics in ribavirin delivery systems, suggesting optimal conditions for effective treatment.
We determined output characteristics (aerosol size and concentration) as a function of time of the Small Particle Aerosol Generator (SPAG-2, 6000 series), which is supplied for use with ribavirin. METHOD: We measured particle size by cascade impactor and aerosol concentration by spectrophotometric analysis. RESULTS: Aerosol concentrations and particle sizes increased with time, especially during prolonged periods of aerosol generation. Aerosolized drug concentrations in a typical administration hood were within 5% of those emitted by the generator unless the hood was placed in a draft, which caused the loss of significant aerosol. Changes in aerosol concentration and total output occurred when either airflows or reservoir volumes were changed. CONCLUSION: Every effort should be made to sustain maximum respirable concentrations by paying close attention to the aerosol delivery system and its operating conditions.
No takes yet. Share an insight, caveat, or question.
Byron et al. (1988) studied this question.