A unipolar stimulus biasing ratio established for the modulation of the 1 kHz tone burst SP separated the Meniere's ears from the non-Meniere's ears with a sensitivity of 85% at a specificity of 80.6% and the difference between groups reached statistical significance (p=0.016). The 1 kHz SP amplitude measurements and the stimulus biasing measurements were superior to the click SP/AP ratio for identifying the Meniere's group. A combination of 1 kHz SP amplitude measurements and SP bias ratio separated the Meniere's ears from the non-Meniere's ears with an accuracy of 85%.
No takes yet. Share an insight, caveat, or question.
Iseli et al. (2010) studied this question.