Validity of the Empatica E4 Wristband to Measure Heart Rate Variability (HRV) Parameters: a Comparison to Electrocardiography (ECG)

ASAngela A. T. SchuurmansPLPeter de LooffKNKarin S. Nijhof

Structured PICO

Does the Empatica E4 wristband accurately measure HR and HRV compared to ECG in traumatized adolescents?

P
Population
15 traumatized adolescents in residential care (yielding 345 recordings)
I
Intervention
Empatica E4 wristband
C
Comparator
VU University Ambulatory Monitoring System (VU-AMS) (ECG)
O
Outcome
Accuracy and predictive value of HR and HRV parameters (SDNN, RMSSD, HF, LF)surrogate

The Empatica E4 wristband demonstrates potential as a valid tool for measuring HR and HRV under non-movement conditions.

Abstract

Wearable monitoring devices are an innovative way to measure heart rate (HR) and heart rate variability (HRV), however, there is still debate about the validity of these wearables. This study aimed to validate the accuracy and predictive value of the Empatica E4 wristband against the VU University Ambulatory Monitoring System (VU-AMS) in a clinical population of traumatized adolescents in residential care. A sample of 345 recordings of both the Empatica E4 wristband and the VU-AMS was derived from a feasibility study that included fifteen participants. They wore both devices during two experimental testing and twelve intervention sessions. We used correlations, cross-correlations, Mann-Whitney tests, difference factors, Bland-Altman plots, and Limits of Agreement to evaluate differences in outcomes between devices. Significant correlations were found between Empatica E4 and VU-AMS recordings for HR, SDNN, RMSSD, and HF recordings. There was a significant difference between the devices for all parameters but HR, although effect sizes were small for SDNN, LF, and HF. For all parameters but RMSSD, testing outcomes of the two devices led to the same conclusions regarding significance. The Empatica E4 wristband provides a new opportunity to measure HRV in an unobtrusive way. Results of this study indicate the potential of the Empatica E4 as a practical and valid tool for research on HR and HRV under non-movement conditions. While more research needs to be conducted, this study could be considered as a first step to support the use of HRV recordings provided by wearables.