Evaluation demonstrates performance of adult resuscitators in extreme temperatures, indicating safety concerns.
Because of the temperature extremes encountered during emergency resuscitation and transport in the field, we sought to evaluate the performance and safety of 10 adult resuscitators (5 permanent units: Hope 4, Laerdal, Lifesaver, Mark 3, and PMR; and 5 disposable units: BagEasy, Code Blue, CPR Bag, DMR, and SPUR) across an operational temperature range of -18°C to 50°C. METHOD: We tested the devices against the American Society for Testing and Materials (ASTM) Standard F-920 and the International Organization for Standardization (ISO) Standard 8382. We tested each resuscitator by using a lung model, the Bio-Tek VT-1 Ventilator Tester. RESULTS: All of the resuscitators met the ventilation requirements for Vr and f (600 mL × 20) and I:E < 1:1 except the SPUR at -18°C. Standards ASTM F-920 and ISO 8382 specify a fractional delivered oxygen concentration (FDO₂) of ≥ 0.85 with attachments and ≥ 0.40 without attachments at oxygen flow of 15 L/min and VE of 7.2 L/min (600 mL × 12). Nine resuscitators met Standards ASTM F-920 and ISO 8382 for FDO₂ with attachments at 21°C and 50°C, but only 3 units (Code Blue, DMR, and PMR) passed at -18°C. At 21°C, the Hope 4 had an FDO2 of 0.77 ± 0.03, which was significantly lower (p < 0.001) than that of the other 9 resuscitators, all of which were ≥ 0.93. Nine resuscitators met the FDO₂ standard without attachments. All 10 resuscitators passed the tests for valve function after contamination with simulated vomitus (at an oxygen flow of 30 L/min) and for backward leakage. At the ventilation pattern recommended by the American Heart Association (AHA) (800 mL × 12) the PMR's mean FDO₂ dropped to 0.86 ± 0.03 because of air leaking into the bag where it attaches to the patient-valve assembly. All 10 resuscitators passed the test for mechanical shock at 21°C and 50°C, but 3 units failed at -18°C. CONCLUSION: We conclude that only the Code Blue and DMR meet the ASTM and ISO standards for operator-powered adult resuscitators across the operational temperature range of -18°C to 50°C.
No takes yet. Share an insight, caveat, or question.
Barnes et al. (1991) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: