This design proposes a generator to reduce power demand in MPI scanners, suggesting significant operational improvements.
The selection-field generator of a Magnetic Particle Imaging (MPI) scanner is typically the system’s largest power consumer. For potential use in intensive care units, it would be highly desirable to operate MPI scanners from a standard wall outlet without dedicated high-power infrastructure. Several concepts for more efficient selection-field generation have been proposed in recent years. In this work, we propose a novel generator design based on an array of soft-iron core coils arranged at an angle around the scanner bore. This geometry reduces overall electrical power demand while enabling gradient strengths of up to 0.5 T m¹. The design was simulated and optimized using magnetostatic simulations in COMSOL Multiphysics. To ensure reliable modeling of the soft-iron cores, prototype coils were built and their magnetic fields were measured, providing experimental validation of the simulation models.
No takes yet. Share an insight, caveat, or question.
Förger et al. (2026) studied this question.