Porous silica particles were prepared by sol-gel method with some modification to get wide-pore particles. These particles were derivatized with N-phenylmaleimide-methylvinylisocyanate (PMI) and styrene by reversible addition fragmentation chain transfer (RAFT) polymerization to prepare N-phenylmaleimide embedded polystyrene (PMP) stationary phases. Narrow bore stainless steel column (100 × 1.8 mm i.d) was packed by slurry packing method. The chromatographic performance of PMP column was evaluated for the separation of synthetic peptides mixture composed of five peptides (Gly-Tyr, Gly-Leu-Tyr, Gly-Gly-Tyr-Arg, Tyr-Ile-Gly-Ser-Arg, Leucine enkephalin) and tryptic digest of human serum albumin (HAS) respectively. Number of theoretical plates as high as 280,000 plates/m were obtained for peptides mixture at optimum elution condition. Separation performance of the developed column was compared with commercial Ascentis Express RP-Amide column and it was observed that separation performance of PMP column was better than commercial column in terms of separation efficiency and resolution.