April 18, 2026BMC Medical Imaging0 citationsOpen Access

Infant and toddler’s preoperative chest imaging on photon-counting detector CT: a paired-sample comparison with energy-integrating detector CT on image quality and radiation dose

XLXiang-Yu LiuRenji HospitalWZWei ZhouRenji HospitalRSRuo‐Yang ShiShanghai Jiao Tong University

Key Points

  • To compare radiation dose and image quality between photon-counting CT and energy-integrating detector CT in infants and toddlers under real-world clinical protocols.
  • Single-center matched cohort study
  • 51 pairs of infants and toddlers underwent preoperative chest CT
  • Prospective PCCT cohort matched with a retrospective EID-CT cohort
  • Dose metrics (CTDIvol, DLP, SSDE) and image quality assessed
  • Subgroup analyses by age, BMI, and water-equivalent diameter conducted.
  • PCCT group had significantly lower radiation dose metrics (P < 0.001)
  • PCCT showed reduced image noise and improved objective image quality metrics (P < 0.001)
  • No statistically significant differences in subjective image quality between PCCT and EID-CT (P > 0.05)
  • Findings remained consistent across age, BMI, and water-equivalent diameter strata.

Abstract

Photon-counting CT (PCCT) may improve dose efficiency compared with energy-integrating detector CT (EID-CT), but evidence remains limited in very young children undergoing non-contrast preoperative chest CT under real-world clinical protocols. To provide a size-aware matched comparison of radiation dose and image quality between PCCT and EID-CT in infants and toddlers (≤ 36 months) undergoing non-contrast preoperative chest CT, and to examine whether the main findings remained directionally consistent across prespecified strata of age, BMI, and water-equivalent diameter (DW). In this single-center matched cohort study, a prospective PCCT cohort and a retrospective EID-CT cohort were propensity score–matched 1:1 (49 pairs). Dose metrics (CTDIvol, DLP, SSDE) and objective and subjective image quality were evaluated, with subgroup analyses stratified by age ( 0.05). Across prespecified age, BMI, and DW strata, PCCT remained associated with lower dose metrics and favorable objective image-quality trends, although these subgroup findings should be interpreted as exploratory. In this matched cohort of infants and toddlers undergoing non-contrast preoperative chest CT, PCCT was associated with substantially lower radiation dose and improved objective image quality compared with EID-CT, and no statistically significant difference was detected in subjective image quality scores between PCCT and EID-CT.