April 20, 2026Journal of the American College of Cardiology596 citations

Incidence and clinical course of thrombus formation on atrial septal defect and patient foramen ovale closure devices in 1,000 consecutive patients

View Full Paper
UKUlrike KrumsdorfSOS. OstermayerKBKai Billinger

Key Result

Thrombus formation after catheter closure of intra-atrial shunts occurred in 20 of 1000 patients, with the Amplatzer device showing a significantly lower incidence (0%) than other devices (p<0.05).

Key Points

  • The study aims to investigate the incidence and clinical implications of thrombus formation after closure of intra-atrial shunts.
  • 1,000 consecutive patients were monitored post-catheter closure of PFO and ASD.
  • Transesophageal echocardiography was performed at four weeks and six months.
  • Additional TEEs were conducted as clinically indicated.
  • Thrombus was found in 5 of 407 ASD patients (1.2%) and in 15 of 593 PFO patients (2.5%).
  • Thrombus resolution occurred in 17 of 20 patients under anticoagulation therapy.
  • Significant predictors of thrombus formation included post-procedure atrial fibrillation and persistent atrial septal aneurysm (p < 0.05).

Study Design

Type

Observational (n=1,000)

Structured PICO

Does the type of closure device affect the incidence of thrombus formation in patients undergoing transcatheter closure of intra-atrial shunts?

P
Population
1,000 consecutive patients undergoing transcatheter closure of patent foramen ovale (PFO) (n = 593) or atrial septal defect (ASD) (n = 407)
I
Intervention
Transcatheter closure of intra-atrial shunts using various devices (CardioSEAL, StarFLEX, PFO-Star, ASDOS, Helex, Amplatzer)
C
Comparator
Comparison among different closure devices
O
Outcome
Incidence, morphology, and clinical course of thrombus formation after catheter closure assessed by transesophageal echocardiographysafety

Thrombus formation on intra-atrial shunt closure devices is generally low (2.0%) and usually resolves with anticoagulation, but the incidence varies significantly depending on the specific device used.

Abstract

OBJECTIVES: The purpose of this study was to investigate the incidence, morphology, and clinical course of thrombus formation after catheter closure of intra-atrial shunts. BACKGROUND: Post-procedure detailed information about thrombotic material on different devices for transcatheter closure is missing. METHOD: A total of 1,000 consecutive patients were investigated after patent foramen ovale (PFO) (n = 593) or atrial septal defect (ASD) (n = 407) closure. Transesophageal echocardiography (TEE) was scheduled after four weeks and six months. Additional TEEs were performed as clinically indicated. RESULTS: Thrombus formation in the left atrium (n = 11), right atrium (n = 6), or both (n = 3) was found in 5 of the 407 (1.2%) ASD patients and in 15 of the 593 (2.5%) PFO patients (p = NS). The thrombus was diagnosed in 14 of 20 patients after four weeks and in 6 of 20 patients later on. The incidence was: 7.1% in the CardioSEAL device (NMT Medical, Boston, Massachusetts); 5.7% in the StarFLEX device (NMT Medical); 6.6% in the PFO-Star device (Applied Biometrics Inc., Burnsville, Minnesota); 3.6% in the ASDOS device (Dr. Ing, Osypka Corp., Grenzach-Wyhlen, Germany); 0.8% in the Helex device (W.L. Gore and Associates, Flagstaff, Arizona); and 0% in the Amplatzer device (AGA Medical Corp., Golden Valley, Minnesota). The difference between the Amplatzer device on one hand and the CardioSEAL device, the StarFLEX device, and the PFO-Star device on the other hand was significant (p < 0.05). A pre-thrombotic disorder as a possible cause of the thrombus was found in two PFO patients. Post-procedure atrial fibrillation (n = 4) and persistent atrial septal aneurysm (n = 4) had been found as significant predictors for thrombus formation (p < 0.05). In 17 of the 20 patients, the thrombus resolved under anticoagulation therapy with heparin or warfarin. In three patients, the thrombus was removed surgically. CONCLUSIONS: The incidence of thrombus formation on closure devices is low. The thrombus usually resolves under anticoagulation therapy.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Krumsdorf et al. (2004) conducted an observational in Patent foramen ovale (PFO) or atrial septal defect (ASD) (n=1,000). Catheter closure devices was evaluated on Thrombus formation. Thrombus formation after catheter closure of intra-atrial shunts occurred in 20 of 1000 patients, with the Amplatzer device showing a significantly lower incidence (0%) than other devices (p<0.05).

synapsesocial.com/papers/69e643e3c3fdccbe27a8aa61https://doi.org/10.1016/j.jacc.2003.10.030
Ask AI
Helpful
Bookmark
Share
View Full Paper