Toward XRF-like chemical sensitivity in the laboratory: first hyperspectral x-ray imaging demonstration using the novel CITIUS detector

VTV. Di TrapaniPTP. ThibaultFAF. Arfelli

Key Points

  • This research aims to demonstrate the use of the CITIUS hyperspectral detector for chemical-specific imaging in laboratory environments.
  • Applied the CITIUS hyperspectral detector to X-ray micro-CT and radiography using a liquid MetalJet source.
  • Analyzed datasets of multi-material test samples with a modified Minimum-Residual Basis Material Decomposition (MR-BMD) algorithm.
  • Focused on capturing sharp energy resolution in the 0.5–0.8 keV range for better material separation.
  • Achieved fine energy resolution and effective material identification, surpassing traditional methods.
  • Successfully separated materials with similar attenuation, such as water and polypropylene.
  • Demonstrated a promising approach toward XRF-like chemical sensitivity in laboratory conditions.

Abstract

Abstract X-ray spectral imaging is an advanced technique that enables material-sensitive imaging by exploiting energy-dependent interactions with matter. In this context, X-ray fluorescence (XRF) represents the benchmark technique. Despite its excellent elemental sensitivity, however, XRF is intrinsically inefficient because it relies on the detection of isotropically emitted secondary fluorescence induced by a focused beam on the sample. This requirement makes the technique difficult to implement with compact laboratory sources and, especially for tomographic applications, highly brilliant synchrotron sources are required. In a laboratory environment, spectral imaging and tomography are typically performed using photon-counting detectors. While this technology offers significant advantages, the limited number of energy thresholds and coarse energy resolution can hamper the separation of materials with similar attenuation properties. Hyperspectral detectors, featuring sub-keV energy resolution and virtually unlimited spectral binning, provide a technological solution to enable high-sensitivity, chemical-specific imaging in the laboratory. Here we present the first application of the novel CITIUS hyperspectral detector to X-ray micro-CT and radiography at the OptImaTo (Optimal Imaging and Tomography) laboratory (Trieste, Italy), based on a liquid MetalJet source (Excillum, Sweden) with a galinstan anode. Using multiple characteristic emission lines (Ga, In, Sn) and a 55 Fe source, a sharp energy resolution in the 0.5–0.8 keV range (full width at half maximum) was found, enabling fine energy binning suitable for advanced quantitative material identification. For the first demonstration, planar and tomographic datasets of two multi-material test samples were analyzed using a newly adapted version of the Minimum-Residual Basis Material Decomposition (MR-BMD) algorithm, optimized for hyperspectral detectors providing tens of energy bins with narrow bandwidths. Results show that the laboratory-based hyperspectral approach combined with MR-BMD enables element-sensitive imaging and, remarkably, separates materials with very similar attenuation, such as water and polypropylene. These results demonstrate accurate material identification and quantification, promisingly approaching XRF-like chemical sensitivity in the laboratory.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Trapani et al. (2026) studied this question.

synapsesocial.com/papers/69eb08ef553a5433e34b39fbhttps://doi.org/10.1038/s41598-026-48904-6
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Quantitative dual-energy micro-CT with a photon-counting detector for material science and non-destructive testing2019 · 31 citations
  2. 2Liquid-metal-jet anode electron-impact x-ray source2003 · 227 citations
  3. 3Laboratory 3D Micro-XRF∕Micro-CT Imaging System2011 · 12 citations
  4. 4A New Method of Interpolation and Smooth Curve Fitting Based on Local Procedures1970 · 2,080 citations
  5. 5Measurements of a dual-energy fast photon counting CdTe detector with integrated charge sharing correction2013 · 63 citations