Answers are grounded in this paper and cite their sources.
Novel transducer demonstrates wireless biochemical sensing in various pH conditions, suggesting efficient long-range applications.
This paper reports on a novel transducer for wireless biochemical sensing. The bilayer transducer consists of a fractal piezoelectric membrane, laminated with a pH-sensitive chemo-mechanical hydrogel. The proposed scheme utilizes the piezoelectric element as a wireless transducer and the phase transition behavior of hydrogels as a sensing element. Notably, the fractal design on the piezoelectric membrane enhances frequency response and linearity by employing a periodically repeated pore architecture. As a basis of the pore, a modified Hilbert space-filling curve is used. By laminating a pH-sensitive hydrogel onto a fractal-porous piezoelectric thin film, the overall transducer curls inward or outward, depending on the environmental pH (e.g., pH = 4, 8, and 12). The curvature exhibits a sensitivity of 10.5°/pH, yielding accordant ultrasound responses under excitation. The measured voltage outputs from an ultrasonic receiver were 0.393, 0.341, and 0.250 mV/cm² for curvature angles of 30°, 60°, and 120°, respectively. Overall pH sensitivity was 0.017 mV/cm²/pH. Ultimately, the biochemical sensing principle using a novel bilayer ultrasound transducer suggests a simple, low-cost, battery-less, and long-range wireless readout system as compared to traditional biochemical sensing.
No takes yet. Share an insight, caveat, or question.
Selvarajan et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: