April 26, 2026Circulation Arrhythmia and Electrophysiology3 citations

Maximizing Lesion Depth With the Lattice-Tip Catheter: A Preclinical Comparison of Single, Double, and Mixed Pulsed-Field and Radiofrequency Applications

View Full Paper
KWKeita WatanabeMSMasaya ShinoharaFYFengyuan Yu

Key Result

Repeating pulsed-field applications twice produced the deepest lesions (6.8±0.3 mm) with the least variation compared to single RF (5.1±0.2 mm, P=0.015) or other ablation strategies.

Key Points

  • The aim is to determine the most effective application of RF and PF to maximize lesion depth during ablation.
  • Sixteen swine underwent endocardial ablation using a lattice-tip catheter.
  • Lesion profiles were created using single and combined RF and PF applications.
  • Analysis focused on lesion depth and variability across different application settings.
  • PF 1-PF 2 produced the deepest mean lesion depth at 6.8 mm (P=0.015 vs RF 1-only).
  • Mean lesion depths did not differ significantly between PF 1-PF 2 and other combination settings (P=0.114, P=0.986).
  • PF 1-PF 2 demonstrated significantly less variability in depth compared to PF 1-only and RF 1-PF 1 (P=0.017, P=0.013).

Structured PICO

Does repeating pulsed-field applications maximize lesion depth and minimize variability compared to single or mixed PF/RF applications in a swine model?

P
Population
16 swine undergoing endocardial ablation in the ventricles
I
Intervention
Two consecutive 4-second pulsed-field applications (PF1-PF2) using a large-footprint lattice-tip catheter
C
Comparator
Single PF (PF1-only), single 5-second saline-irrigated temperature-controlled RF (RF1-only), PF followed by RF (PF1-RF1), and RF followed by PF (RF1-PF1)
O
Outcome
Lesion depth and variabilitysurrogate

Repeating pulsed-field applications twice with a lattice-tip catheter optimizes lesion depth and consistency compared to single or mixed PF/RF strategies in a preclinical model.

Main Result

Absolute Event Rate: 6.8% vs 5.1%

p-value: p=0.015

Abstract

BACKGROUND: Combining radiofrequency (RF) and pulsed-field (PF) applications in various combinations, or repeating PF applications, has been reported to augment lesion size and depth. We compared lesion dimensions using single, double, and combined applications of PF and RF for the large-footprint lattice-tip catheter. METHODS: Sixteen swine underwent endocardial ablation using the lattice-tip catheter. Four-second PF and 5-second, saline-irrigated, temperature-controlled RF energy were delivered in the ventricles using 1 of 5 different settings: single applications of PF (PF 1 -only), single RF (RF 1 -only), and 3 sequential settings: PF 1 -PF 2 (2 consecutive PF applications at the same location), PF 1 -RF 1 (PF followed by RF), and RF 1 -PF 1 (RF followed by PF). RESULTS: A total of 102 lesions (69 left and 33 right ventricles) were analyzed. Mean lesion depths were 6.8±0.3 mm (PF 1 -PF 2 ), 6.6±0.3 mm (RF 1 -PF 1 ), 5.7±0.2 mm (PF 1 -only), 5.0±0.3 mm (PF 1 -RF 1 ), and 5.1±0.2 mm (RF 1 -only). Mean lesion depths did not differ significantly between PF 1 -PF 2 and either PF 1 -only or RF 1 -PF 1 ( P =0.114 and P =0.986, respectively). However, the variability in depth was significantly smaller in the PF 1 -PF 2 group compared with both PF 1 -only and RF 1 -PF 1 ( P =0.017 and P =0.013, respectively). Only the PF 1 -PF 2 setting produced significantly deeper lesions compared with the RF 1 -only and PF 1 -RF 1 settings ( P =0.015 and P =0.012, respectively). RF lesions exhibited a prominent dark hemorrhagic core and a thin pale rim; PF lesions appeared uniformly pale; and combination lesions demonstrated a central dark zone surrounded by a pale rim of variable thickness. CONCLUSIONS: Repeating PF twice created the optimal combination of greatest depth and least variation compared with all other strategies. RF 1 -PF 1 resulted in similar depth, but required toggling between settings and risked thermal adverse events.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Watanabe et al. (2026) studied Endocardial ablation (preclinical) (n=16). Sequential pulsed-field applications (PF1-PF2) vs. Single RF, single PF, or mixed PF/RF applications was evaluated on Mean lesion depth (p=0.015). Repeating pulsed-field applications twice produced the deepest lesions (6.8±0.3 mm) with the least variation compared to single RF (5.1±0.2 mm, P=0.015) or other ablation strategies.

synapsesocial.com/papers/69edad094a46254e215b4b14https://doi.org/10.1161/circep.125.014799
Ask AI
Helpful
Bookmark
Share
View Full Paper