Key result
Analysis software influences coronary calcium volume scores more than CT scanner type.
Why the study?
Repeat coronary artery calcium scoring should provide robust results with low inter-scanner variability to allow meaningful comparison for risk stratification and monitoring.
Does the use of different CT scanners (64-MSCT vs 64-DSCT) affect the variability of coronary artery calcium scoring in patients?
Observational (n=30)
Does the use of different CT scanners (64-MSCT vs 64-DSCT) affect the variability of coronary artery calcium scoring in patients?
Inter-scanner variability for CAC scoring is reasonable, especially for Agatston scores <1000, but different software solutions significantly influence volume scores.
AIMS: Coronary artery calcium (CAC) scoring has emerged as a tool for risk stratification and potentially for monitoring response to risk factor modification. Therefore, repeat measurements should provide robust results and low inter-scanner variability for allowing meaningful comparison. The purpose of this study was to investigate inter-scanner variability of CAC for Agatston, volume, and mass scores by head-to-head comparison using two different cardiac computed tomography scanners: 64-detector multislice CT (MSCT) and 64-slice dual-source CT (DSCT). METHODS AND RESULTS: Thirty patients underwent CAC measurements on both 64-MSCT (GE LightSpeed XT scanner: 120 kV, 70 mAs, 2.5 mm slices) and 64-DSCT (Siemens Somatom Definition: 120 kV, 80 mAs, 3 mm slices) within <100 days (0-97). Retrospective intra-scan comparison revealed an excellent correlation. The excellent intra-scan (inter-observer) agreement was documented by narrow limits of agreement and a correlation coefficient of variation (COV) of r ≥ 0.99 (P < 0.001) for all CAC scores with a low COV for both scanners (64-MSCT/64-DSCT), i.e. Agatston (2.0/2.1%), mass (3.0/2.0%), and volume (4.7/3.9%). Inter-scanner comparison revealed larger Bland-Altman (BA) limits of agreement, despite high correlation (r ≥ 0.97) for all scores, with COV at 15.1, 21.6, and 44.9% for Agatston, mass, and volume scores. The largest BA limits were observed for volume scores (-1552.8 to 574.2), which was massively improved (-241.0 to 300.4, COV 11.5%) after reanalysing the 64-DSCT scans (Siemens) with GE software/workstation (while Siemens software/workstation does not allow cross-vendor analysis). Phantom measurements confirmed overestimation of volume scores by 'syngo Ca-Scoring' (Siemens) software which should therefore be reviewed (vendor has been notified). CONCLUSION: Intra- and inter-scan agreement of CAC measurement in a given data set is excellent. Inter-scanner variability is reasonable, particularly for Agatston units in the clinically most relevant range <1000. The use of different software solutions has a greater influence particularly on volume scores than the use of different scanner types.
No takes yet. Share an insight, caveat, or question.
Ghadri et al. (2011) conducted an observational in Coronary artery calcium (n=30). 64-multidetector computed tomography (64-MSCT) vs. 64-slice dual-source computed tomography (64-DSCT) was evaluated on Inter-scanner variability of CAC for Agatston, volume, and mass scores. Coronary artery calcium scoring showed high inter-scanner correlation (r ≥ 0.97) between 64-MSCT and 64-DSCT, though volume scores were heavily influenced by different software solutions.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: