PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
November 1, 1993Clinical Physiology37 citations

Reproducibility of cardiac volume measurements including left ventricular mass determined by MRI,

View Full Paper
JMJens MøgelvangResonance Research (United States)KLK LindvigUniversity of Copenhagen
Lars Søndergaard
Lars SøndergaardInterventional / Structural Cardiology

Structured PICO

Does magnetic resonance imaging (MRI) using a contiguous multislice technique provide reproducible cardiac volume measurements compared to simplified MRI and 2D echocardiography?

P
Population
30 patients undergoing cardiac volume measurements
I
Intervention
Magnetic resonance imaging (MRI) using a contiguous multislice technique
C
Comparator
Method using only one long axis and four short axis slices, and two-dimensional echocardiography (in a subset of 13 patients)
O
Outcome
Intra- as well as interobserver and interstudy variation of cardiac volume measurements including left ventricular masssurrogate

While contiguous multislice MRI provides reproducible measurements of left ventricular mass, it cannot be used interchangeably with simplified MRI or 2D echocardiography due to wide limits of agreement.

Abstract

Cardiac volumes including left ventricular mass were determined by magnetic resonance imaging (MRI) in 30 patients using a contiguous multislice technique. Eleven patients were studied on two different days. Intra- as well as interobserver and interstudy variation were evaluated. For volume measurements the standard deviation of differences (SDD) was rather constant, independent of the size of the measured volume indicating lower percentual variability in measuring large volumes. Lowest variability was noticed for left ventricular myocardial mass estimates between observers as well as between studies. In 13 patients left ventricular mass was estimated by this technique as well as by a method using only one long axis and four short axis slices and by two-dimensional echocardiography. Although highly correlated, limits of agreement between the methods were too wide to accept interchangeable clinical use of the methods.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Møgelvang et al. (1993) studied this question.

synapsesocial.com/papers/69f7e3eaa00cd75688a510achttps://doi.org/10.1111/j.1475-097x.1993.tb00474.x
Ask AI
Helpful
Bookmark
Share
View Full Paper