BACKGROUND: (NG) is limited. OBJECTIVES: To review the performance and operational characteristics of NG POCTs for diagnosis of urogenital gonorrhoea. METHODS: We compiled and synthesised findings from two separate systematic reviews which included evaluations published until August 2015. RESULTS: Six tests were included: five were immunochromatographic tests (ICTs) or optical immunoassay (OIAs) based on antigen detection; with 5-7 steps and results in 25-40 min, and one (GeneXpert CT/NG) was a 'near-patient test' based on nucleic acid amplification technique (NAAT); with three steps, electricity required, and results in 90 min. When compared with laboratory-based NAATs as the reference tests, sensitivities of ICT and OIA-based POCTs ranged from 12.5% to 70% when cervical/vaginal swabs were tested. Specificities ranged from 89% to 99.8%. The near-patient NAAT had sensitivities of >95% and specificities of >99.8% consistently across all specimen types (urine, cervical and vaginal swabs). CONCLUSIONS: Based on a limited number of evaluations, antigen detection POCTs for NG lacked sufficient sensitivity to be used for screening. A near-patient NAAT has acceptable performance, only involved a few steps, but needs electricity, a temperature-controlled environment and has a 90 min run time. To achieve wider scale up of NG POCTs, we need strong evidence of cost-effectiveness, which should inform guidelines and ultimately increase test development, demand and reduce costs.