BACKGROUND: To be able to compare continuous glucose monitoring (CGM) systems, they have to be worn in parallel by the same subjects. This study evaluated the performance and usability of three different CGM systems in direct comparison. METHOD: In this open, prospective study at two sites, 54 patients with diabetes wore three CGM systems each (Dexcom G5™ Mobile CGM system DG5, Guardian™ Connect system GC, and a Roche CGM system RCGM) in parallel for 6 or 7 days in a mixed inpatient and outpatient setting. Capillary comparison measurements were performed using a self-monitoring of blood glucose (SMBG) system. During study site visits, glucose excursions were induced. Performance of the systems was evaluated by calculating mean absolute relative differences (MARD, calculated as absolute differences for glucose concentrations 99% of values within zones A and B of the consensus error grid. Overall, all CGM systems showed a small negative bias compared to SMBG. Usability of the systems differed regarding patch adhesion rate, failure rate, and patient rating. Most patients preferred GC, but in general all systems were rated positively. CONCLUSION: All three CGM systems showed similar overall accuracy in this direct comparison, but small differences were observed with regard to specific glucose ranges and usability aspects.