Answers are grounded in this paper and cite their sources.
high-exertion exercise, electrical stimulation (such as iontophoresis requiring electrical power), or chemical stimulation (such as by delivering pilocarpine or carbachol inside skin), to extract sweat under low-perspiring conditions such as at rest. Here, we present a continuous sweat lactate monitoring platform combining a hydrogel for osmotic sweat extraction, with a paper microfluidic channel for facilitating sweat transport and management, a screen-printed electrochemical lactate sensor, and a custom-built wireless wearable potentiostat system. Osmosis enables zero-electrical power sweat extraction at rest, while continuous evaporation at the end of a paper channel allows long-term sensing from fresh sweat. The positioning of the lactate sensors provides near-instantaneous sensing at low sweat volume, and the custom-designed potentiostat supports continuous monitoring with ultra-low power consumption. For a proof of concept, the prototype system was evaluated for continuous measurement of sweat lactate across a range of physiological activities with changing lactate concentrations and sweat rates: for 2 h at the resting state, 1 h during medium-intensity exercise, and 30 min during high-intensity exercise. Overall, this wearable system holds the potential of providing comprehensive and long-term continuous analysis of sweat lactate trends in the human body during rest and under exercising conditions.
No takes yet. Share an insight, caveat, or question.
Saha et al. (2022) studied this question.