Investigates the effect of graphite coating on thermal diffusivity in silicon nitride substrates, implying importance for accurate measurements.
The influence of graphite coating on the measurement of thermal diffusivity by the flash method was systematically investigated using thin, high-thermal-conductivity silicon nitride (Si 3 N 4 ) substrates (≈90 W/(m·K) and 120 W/(m·K)) with thicknesses of 2.0 mm, 0.32 mm, and 0.25 mm. While coating had negligible influence for 2.0 mm samples, the measured diffusivity decreased with increasing coating weight for 0.32 mm and 0.25 mm samples, with stronger effects for thinner and more thermally conductive substrates. The results highlight the need to strictly control graphite-coating amount for thin, high-conductivity materials, and the observed trend was quantitatively explained using a laminated structure model accounting for through-thickness heat transfer.
No takes yet. Share an insight, caveat, or question.
Zhou et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: