Key result
Cine MRI with 3D contrast MRA achieves 100% diagnostic accuracy for APVs and ASDs.
Why the study?
Precise anatomic definition of anomalous pulmonary and systemic veins and the atrial septum is required for surgical correction of ASDs, but cardiac catheterization and TEE did not provide complete information in these patients.
Does fast cine MRI with 3D contrast-enhanced MRA provide accurate anatomic definition of APVs and ASDs in patients with inconclusive catheterization or TEE?
Observational (n=20)
Does fast cine MRI with 3D contrast-enhanced MRA provide accurate anatomic definition of APVs and ASDs in patients with inconclusive catheterization or TEE?
Fast cine MRI with 3D contrast-enhanced MRA is highly accurate for diagnosing anomalous pulmonary veins and atrial septal defects, particularly when standard imaging modalities are inconclusive.
OBJECTIVES: The purpose of our study was to evaluate patients with suspected anomalous pulmonary veins (APVs) and atrial septal defects (ASDs) using fast cine magnetic resonance imaging (MRI) and ultrafast three-dimensional magnetic resonance angiography (MRA). BACKGROUND: Precise anatomic definition of anomalous pulmonary and systemic veins, and the atrial septum are prerequisites for surgical correction of ASDs. Cardiac catheterization and transesophageal echocardiography (TEE) are currently used to diagnose APVs, but did not provide complete information in our patients. METHODS: Twenty consecutive patients with suspected APVs were studied by MRA after inconclusive assessment by catheterization, TEE or both. The MRI images were acquired with a fast cine sequence and a novel ultrafast three-dimensional sequence before and after contrast injection. RESULTS: Partial anomalous pulmonary venous drainage was demonstrated in 16 of 20 patients and was excluded in four patients. Magnetic resonance imaging correctly diagnosed APVs and ASDs in all patients (100%) who underwent surgery. For the diagnosis of APVs, the MRI and catheterization results agreed in 74% of patients and the MRI and TEE agreed in 75% of patients. For ASDs, MRI agreed with catheterization and TEE in 53% and 83% of patients, respectively. CONCLUSIONS: Fast cine MRI with three-dimensional contrast-enhanced MRA provides rapid and comprehensive anatomic definition of APVs and ASDs in patients with adult congenital heart disease in a single examination.
No takes yet. Share an insight, caveat, or question.
Ferrari et al. (2001) conducted an observational in Suspected anomalous pulmonary veins (APVs) and atrial septal defects (ASDs) (n=20). Fast cine MRI and ultrafast three-dimensional contrast-enhanced MRA vs. Cardiac catheterization and transesophageal echocardiography (TEE) was evaluated on Correct diagnosis of APVs and ASDs confirmed by surgery. Fast cine MRI with three-dimensional contrast-enhanced MRA correctly diagnosed anomalous pulmonary veins and atrial septal defects in 100% of patients who underwent surgery.