April 1, 2000Diabetes Care141 citationsOpen Access

Evaluation of an electrochemical sensor for measuring blood ketones.

HBHenry ByrneUniversity College DublinKTK.L. TieszenSalford Royal HospitalSHS. HollisRoyal United Hospital Bath NHS Trust

Key Points

Key points are not available for this paper at this time.

Abstract

OBJECTIVE: To evaluate the performance of a hand-held ketone sensor that is able to measure blood beta-hydroxybutyrate (beta-HBA) concentrations within 30 s in patients with diabetic ketoacidosis (DKA) and patients who attend a weight management clinic. RESEARCH DESIGN AND METHODS: Two groups of patients were studied: 19 patients admitted with DKA and 156 patients attending a weight management clinic. Paired capillary and venous whole blood samples were measured using the ketone sensor and also using an enzymatic laboratory reference method. RESULTS: The ketone sensor accurately measured beta-HBA concentrations in patients with DKA (limits of agreement -0.9 to + 1.0 mmol/l) or starvation-induced ketonemia (limits of agreement -0.5 to +0.5 mmol/l). CONCLUSIONS: This ketone sensor accurately measures whole blood beta-HBA concentrations within 30 s.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Byrne et al. (2000) studied this question.

synapsesocial.com/papers/69ffa423581c6e761e778167https://doi.org/10.2337/diacare.23.4.500