Why the study?
The initial, multi-center commercial implant experience of the AVEIR DR dual-chamber leadless pacemaker system had yet to be evaluated.
Is the AVEIR DR dual-chamber leadless pacemaker safe and efficient in patients indicated for dual-chamber pacing in a real-world commercial setting?
Is the AVEIR DR dual-chamber leadless pacemaker safe and efficient in patients indicated for dual-chamber pacing in a real-world commercial setting?
The initial commercial experience of the AVEIR DR dual-chamber leadless pacemaker demonstrates safe and efficient implantation with a 99% complication-free rate at 30 days.
INTRODUCTION: The AVEIR DR pacemaker (Abbott) is a dual-chamber leadless pacemaker (LP) system that includes right atrial and right ventricular helix-fixation LPs (ALP, VLP), with atrioventricular synchrony maintained via wireless communication between the two devices. The initial, multi-center commercial implant experience of the AVEIR DR dual-chamber LP system has yet to be evaluated. METHODS: Patients indicated for dual-chamber pacing after US commercial release were consecutively included. Implant procedural characteristics were evaluated, and electrical parameters were measured during pre-fixation mapping, post-fixation tether mode, after LP release, and before patient discharge. Any acute procedure- or device-related complications within 30 days were noted. RESULTS: Patients were implanted with AVEIR DR per standard practice (N = 175 at 8 centers; 72 ± 11 years; 60% male; 62% sinus node dysfunction, 37% AV block; 83% de novo implants). ALPs were implanted predominantly in the base of the right atrial appendage (82%), VLPs in the mid-to-apical right ventricular septum (94%). Pre-fixation mapping allowed repositioning to be avoided in 95% of ALPs and VLPs. The total procedure duration was 64 ± 33 min, from initial incision to final suture, with a cumulative fluoroscopy duration of 13 ± 8 min. In both LPs, capture thresholds and sensed amplitudes improved significantly before patient discharge (ALP 1.2 ± 1.0 V, 2.1 ± 1.1 mV; VLP 0.8 ± 0.6 V, 9.5 ± 3.8 mV). At 30 days post-implant, 99% of patients were complication-free. CONCLUSION: The initial commercial experience of the helix-fixation, dual-chamber LP system demonstrated safe and efficient implantation with clinically acceptable electrical metrics and minimal acute complications.
No takes yet. Share an insight, caveat, or question.
Hadadi et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: