Optoacoustic tomography provides a unique possibility for ultra-high-speed 3-D imaging by acquiring complete volumetric datasets from interrogation of tissue by a single nanosecond-duration laser pulse. Yet, similarly to ultrasound, optoacoustics is a time-resolved imaging method, thus, fast 3-D imaging implies real-time acquisition and processing of high speed data from hundreds of detectors simultaneously, which presents significant technological challenges. Herein we present a highly efficient graphical processing unit (GPU) framework for real-time reconstruction and visualization of 3-D tomographic optoacoustic data. By utilizing a newly developed 3-D optoacoustic scanner, which simultaneously acquires signals with a handheld 256-element spherical ultrasonic array system, we further demonstrate tracking of deep tissue human vasculature rendered at a rate of 10 volumetric frames per second. The flexibility provided by the handheld hardware design, combined with the real-time operation, makes the developed platform highly usable for both clinical imaging practice and small animal research applications.
No takes yet. Share an insight, caveat, or question.
Deán‐Ben et al. (2013) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: