PURPOSE: To compare different new-generation biometric formulas and ray-tracing for small-aperture intraocular lens (IOL) (IC-8; Acufocus, Inc) implantation in patients undergoing cataract and refractive lens exchange surgery with highly irregular corneas. METHODS: This monocenter study included 17 eyes of 17 patients with highly irregular corneas of different genesis. Biometric and topographic corneal data were assessed using the IOLMaster 700 (Carl Zeiss Meditec) and Pentacam (Oculus Optkigeräte GmbH). Prediction and absolute error were compared after 3 months based on manifest refraction. Furthermore, change of total corneal refractive power in different corneal pathologies was also evaluated. For IOL power calculation, three fourth-generation IOL formulas were compared (Haigis, SRK-T, and Barrett Universal II). The dataset was then checked against ray-tracing and analyzed to improve prediction error in these highly irregular corneas. RESULTS: All patients showed an improvement in visual acuity postoperatively with a mean spherical equivalent of -1.22 ± 0.49 diopters (D). Overall comparison of the three formulas showed the Haigis formula to be superior in terms of the smallest deviation of predictive and absolute error. IOL calculations with ray-tracing were possible in all eyes, but showed inaccurate results with keratometric values of 48.00 D and greater. CONCLUSIONS: .