BACKGROUND: X-ray and particle radiation therapy planning requires accurate estimation of local electron density within the patient body to calculate dose delivery to tumour regions. We evaluate the feasibility and accuracy of electron density measurement using dual-layer computed tomography (DLCT), a recently introduced dual-energy CT technique. METHODS: Two calibration phantoms were scanned with DLCT and virtual monoenergetic images (VMIs) at 50 keV and 200 keV were generated. We investigated two approaches to obtain relative electron densities from these VMIs: to fit an analytic interaction cross-sectional model and to empirically calibrate a conversion function with one of the phantoms. Knowledge of the emitted x-ray spectrum was not required for the presented work. RESULTS: = 0.348, 0.167), suggesting that the methods are reproducible. Further, we applied these methods to routine clinical data. CONCLUSIONS: Our study shows a high validity of electron density estimation based on DLCT, which has potential to improve the procedure and accuracy of measuring electron density in clinical practice.