This is the authors' abstract. We don't add key points for this paper.
This paper describes a new multielectrode microprobe which utilizes integrated-circuit fabrication techniques to overcome many of the problems associated with conventional microelectrodes. The probe structure consists of an array of gold electrodes which are supported on a silicon carrier and which project beyond the carrier for a distance of about 50 μ to allow a close approach to active neurons. These electrodes are covered with a thin (0.4-μ) layer of silicon dioxide which is selectively removed at the electrode tips using high-resolution photoengraving techniques to define the recording areas precisely. The processing sequence described permits any two-dimensional electrode array to be realized. Interelectrode spacings can be accurately controlled in the range from 10 to 20 μ or larger, and electrode-tip diameters can be as small as 2 μ.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Wise et al. (1970) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: