values, which was 108.8 ± 1.4%. OAR indices were also evaluated for Ethos plans using Radiation Therapy Oncology Group report 0415 as a guide and were found to be comparable to each other and the Eclipse plans. While all Ethos plans were comparable, we found that, in general, the Ethos 12-field IMRT plans met most of the dosimetric goals for treatment. Also, Ethos IOE consistently generated dosimetrically hotter VMAT plans versus IMRT plans. On average, Ethos TPS took 13 min to generate 2-full arc VMAT plans, compared to 5 min for 12-field IMRT plans. Varian Ethos TPS can generate multiple treatment plans in an efficient time frame and the quality of the plans could be deemed clinically acceptable when compared to manually generated treatment plans.