Why the study?
Questions have emerged regarding the adequacy of current reprocessing practices for preventing duodenoscopes from transmitting CRE and related MDROs.
Does ethylene oxide gas sterilisation terminate multidrug-resistant bacterial outbreaks linked to duodenoscopes?
Does ethylene oxide gas sterilisation terminate multidrug-resistant bacterial outbreaks linked to duodenoscopes?
Ethylene oxide gas sterilisation may be an effective measure to terminate multidrug-resistant bacterial outbreaks linked to duodenoscopes when standard high-level disinfection fails.
Ethylene oxide sterilisation may terminate duodenoscopy-related MDR outbreaks after failed disinfection; extends case evidence but requires prospective trials.
BACKGROUND: (CRE) and related multidrug-resistant organisms (MDROs), reasonable questions about the adequacy of current practices for reprocessing duodenoscopes have emerged. OBJECTIVES: To review and evaluate the adequacy of current reprocessing practices for preventing duodenoscopes from transmitting CRE and related MDROs. METHODS: The MEDLINE/PubMed database was searched to identify published cases associating confirmed (or suspected) infections of CRE or a related MDRO with exposure to a duodenoscope since 2012, when duodenoscopes became a recognised risk factor for the transmission of CRE. The Internet was also searched to identify news articles and other reports documenting eligible cases occurring during this same timeframe but not identified during the MEDLINE database's search. The Food and Drug Administration's (FDA) medical device database was queried to identify regulatory reports describing these same types of cases, also recorded since 2012. The clinical and reprocessing details of each eligible case were reviewed to identify (when possible): (a) the reprocessing method (e.g., high-level disinfection) performed at the time of the infections, (b) whether the facility's compliance with the manufacturer's reprocessing instructions was confirmed, and (c) the measure(s) or corrective action(s) the facility implemented to prevent additional multidrug-resistant infections. RESULTS: gene as the pathogen. For 12 of these 23 cases, it was reported or implied that the duodenoscope was being high-level disinfected at the time of the infections, consistent with published guidelines. For the remaining 11 cases, the associated report(s) did not clearly identify how the duodenoscope was being reprocessed at the time of the infections (although it may be reasonably concluded that at least some, if not all, of these 11 cases involved high-level disinfection).Further, eight of the 23 cases reported the duodenoscope was being reprocessed in accordance with the manufacturer's instructions for use (and professional guidelines) at the time of the infections. Seven of the cases discussed the design of the duodenoscope (eg, the forceps elevator mechanism) in the context of reprocessing and the infections. Three of the cases identified one or more reprocessing lapses, including inadequate cleaning, delayed reprocessing and improper drying and/or storage of the duodenoscope. Most of these 23 cases were associated with exposure to a duodenoscope model featuring a sealed elevator-wire channel. Six of the cases reported adopting (or in one case supplementing high-level disinfection with) ethylene oxide (EO) gas sterilisation of the duodenoscope, with at least three reporting this measure terminated the outbreak. Other measures adopted to prevent additional infections included removing the implicated duodenoscope from use, re-training staff about proper cleaning, microbiological culturing of the duodenoscope and returning the duodenoscope to the manufacturer for evaluation, maintenance and/or repair. CONCLUSIONS: devices requiring sterilisation is currently unresolved.
No takes yet. Share an insight, caveat, or question.
Lawrence F. Muscarella (2019) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: