August 11, 2021Science Robotics149 citations

Soft robotic manipulator for intraoperative MRI-guided transoral laser microsurgery

View Full Paper
GFGe FangMCMarco C. K. ChowJHJustin Ho

Key Points

  • The research aims to develop a soft robotic manipulator for improving laser microsurgery precision under MRI guidance.
  • Proposed a miniaturized MRI-compatible soft robotic system with five degrees of freedom.
  • Conducted ex vivo tissue ablation and cadaveric trials while monitoring thermal diffusion using MR thermometry.
  • Validated robot operation through laser path-following tests to ensure steering accuracy.
  • Submillimeter laser steering accuracy achieved with a mean error < 0.20 mm.
  • MRI compatibility testing showed zero observable image artifacts during robot operation.
  • Thermal monitoring combined with laser ablation ensured precise control of resection margins.

Abstract

Magnetic resonance (MR) imaging (MRI) provides compelling features for the guidance of interventional procedures, including high-contrast soft tissue imaging, detailed visualization of physiological changes, and thermometry. Laser-based tumor ablation stands to benefit greatly from MRI guidance because 3D resection margins alongside thermal distributions can be evaluated in real time to protect critical structures while ensuring adequate resection margins. However, few studies have investigated the use of projection-based lasers like those for transoral laser microsurgery, potentially because dexterous laser steering is required at the ablation site, raising substantial challenges in the confined MRI bore and its strong magnetic field. Here, we propose an MR-safe soft robotic system for MRI-guided transoral laser microsurgery. Owing to its miniature size (Ø12 × 100 mm), inherent compliance, and five degrees of freedom, the soft robot ensures zero electromagnetic interference with MRI and enables safe and dexterous operation within the confined oral and pharyngeal cavities. The laser manipulator is rapidly fabricated with hybrid soft and hard structures and is powered by microvolume (<0.004 milliter) fluid flow to enable laser steering with enhanced stiffness and lowered hysteresis. A learning-based controller accommodates the inherent nonlinear robot actuation, which was validated with laser path-following tests. Submillimeter laser steering accuracy was demonstrated with a mean error < 0.20 mm. MRI compatibility testing demonstrated zero observable image artifacts during robot operation. Ex vivo tissue ablation and a cadaveric head-and-neck trial were carried out under MRI, where we employed MR thermometry to monitor the tissue ablation margin and thermal diffusion intraoperatively.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Fang et al. (2021) studied this question.

synapsesocial.com/papers/6a0311a1a59fd503299e25b7https://doi.org/10.1126/scirobotics.abg5575
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Interaction Forces of Soft Fiber Reinforced Bending Actuators2016 · 200 citations
  2. 2MR thermometry‐based feedback control of laser interstitial thermal therapy at 980 nm2004 · 219 citations
  3. 3MRI Guided Surgical Robot2001 · 46 citations
  4. 4A shifting paradigm for patients with head and neck cancer: transoral robotic surgery (TORS).2010 · 41 citations