BACKGROUND: The formation of fluid-filled microvacuoles, termed glistenings, is a common complication of intraocular lenses (IOLs) made from hydrophobic acrylate. Using our well-established in-vitro laboratory method, we evaluated a new IOL material's resistance to glistening formation. METHODS: ) was evaluated in five sections of each IOL. The results for each model were compared and rated on a modified Miyata Scale for grading glistening severity. RESULTS: . Mean value of the Eyecryl ASHFY600 IOL, using the Miyata Scale, was Zero. CONCLUSION: In this in-vitro laboratory study, the new hydrophobic acrylic IOL showed a high resistance to microvacuole formation. Results from this in-vitro study suggest that glistening numbers will be low in clinical use in the Eyecryl ASHFY600.