BACKGROUND/PURPOSE: study was to evaluate fatigue resistance of dental fixtures in three different types of fixture/abutment finishing line. MATERIALS AND METHODS: Transmucosal dental implants, with or without ferrulized neck, underwent fatigue tests (static and dynamic load) using the following standard protocol: UNI EN ISO 14801:2016. Two types of loading devices (screw- or cement-retained restoration) were also tested, and fatigue cycle tests were run to failure. Data of static and dynamic load tests were analyzed by proper statistical methods. RESULTS: type (Implant Level with ferrulized neck and Screw-retained crown, 362.8 N). No fracture of the titanium-base was registered in the tested specimens during the static loadings. Permanent deformations of the materials were observed. CONCLUSION: and AL groups (p-values < 0.01): much higher life of one-half order of magnitude.