Answers are grounded in this paper and cite their sources.
Purpose: To report the clinical outcomes of ICL implantation for the treatment of myopia in American patients at a single center with central internal anterior chamber depth (ACD) less than the FDA-approved 3.0 mm. Patients and Results: Of the 231 eyes with ACD < 3.0 mm, 0.4% (1/231) required intraocular surgical intervention including ICL explant, exchange, or toric lens repositioning at some point in the postoperative period or were clinically judged to have a vault requiring an exchange compared to 3.7% (26/712) eyes with ACD ≥ 3.0 mm (p = 0.0103). Average postoperative vault for the eyes with ACD < 3.0 mm was 437.22 ± 191.93 microns compared to 560.73 ± 193.65 in eyes with ACD ≥ 3.0 mm (p < 0.001). IOP and visual acuity at 1 month were similar between the groups. Conclusion: Outside the US, the lower limit of ACD is 2.8 mm for myopia correction. This single center study demonstrates that ICL surgery in American eyes with ACDs < 3.0 mm may also be safe and result in excellent outcomes with comparable rates of exchange or enhancement to eyes meeting the FDA guidelines.
No takes yet. Share an insight, caveat, or question.
Hirabayashi et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: