Answers are grounded in this paper and cite their sources.
Assesses renalase biomarker variability in chronic kidney disease, indicating the need for standardization.
BACKGROUND: Renalase is a promising biomarker for kidney disease, but published levels vary widely between studies. We hypothesised that variability in commercial enzyme-linked immunosorbent assays (ELISAs) kits and matrix effects (serum vs plasma) drive these inconsistencies. METHODS: Paired serum and plasma samples from 56 participants (28 chronic kidney disease (CKD) stages 2-5, 28 healthy controls) were tested using three commercial renalase ELISAs (BTLAB, Cloud-Clone, EIAab). We assessed intra-assay precision, inter-assay agreement (Spearman's rank correlation and Bland-Altman analysis on log10-transformed values), matrix effects, and associations with estimated glomerular filtration rate (eGFR). Diagnostic performance was evaluated by Receiver operating characteristic (ROC) analysis. Cloud-Clone and EIAab produced divergent results and strong matrix-dependent biases. CONCLUSIONS: Observed variability among commercial ELISA platforms may compromise comparability between studies. Harmonisation, standardised reference materials, and cross-validation are necessary before renalase assays can be used reliably in clinical practice.
No takes yet. Share an insight, caveat, or question.
Serwin et al. (2026) studied this question.