May 14, 2026The Journal of the Acoustical Society of America

Primary acoustic measurement using the acousto-optic method in a Fabry-Pérot cavity

View Full Paper
Ask AI
Bookmark
Share

Authors

BRBenjamin J. ReschovskyRARichard A. AllenSFSteven E. Fick

Discussion

Loading...

Member takes

Overview

Randomized trial compares acoustic pressure measurement methods in low-uncertainty settings, suggesting acousto-optic viability.

Key Points

  • The aim is to enhance primary acoustic measurement sensitivity using the acousto-optic method in a Fabry-Pérot cavity.
  • Comparison of acoustic pressure measured by a standard microphone and the acousto-optic method.
  • Utilized a high-finesse Fabry-Pérot cavity and acoustic resonator at 1 kHz.
  • Measured uncertainty comparable to existing methods.
  • Demonstrated agreement in measurement at the percent level.
  • Confirmed the sensitivity enhancements using the acousto-optic method.
  • Indicated potential improvements and viability as an alternative measurement method.

Cite This Study

Reschovsky et al. (2025) studied this question.