Key result
Transfer function model accurately reconstructs internal radial pulse pressure waveforms from external robotic tonometry.
Why the study?
To develop a mathematical transfer function model to accurately estimate and rebuild internal arterial pulse pressure waveforms using non-invasive arterial tonometry measurements.
Does a transfer function model accurately reconstruct internal radial pulse pressure waveforms from external tonometry measurements in an in vitro wrist model?
Does a transfer function model accurately reconstruct internal radial pulse pressure waveforms from external tonometry measurements in an in vitro wrist model?
A novel transfer function model accurately reconstructed internal radial pulse pressure waveforms from non-invasive robotic tonometry measurements in an in vitro setting.
May enable non-invasive waveform estimation; leaves open human validation before clinical use.
The primary goal of this study is to develop a mathematical model that can establish a transfer function relationship between the "external" pulse pressures measured by a tonometer and the "internal" pulse pressure in the artery. The purpose of the model is to accurately estimate and rebuild the internal pulse pressure waveforms using arterial tonometry measurements. To develop and validate a model without human subjects and operators for consistency, this study employs a radial pulse generation system, a robotic tonometry system, and a write model with an artificial skin and vessel. A transfer function model is developed using the results of the pulse testing and the mechanical characterization testing of the skin and vessel. To evaluate the model, the pulse waveforms are first reconstructed for various reference pulses using the model with tonometry data. They are then compared with pulse waveforms acquired by internal measurement (by the built-in pressure sensor in the vessel) the external measurement (the on-skin measurement by the robotic tonometry system). The results show that the model-produced pulse waveforms coinciding well with the internal pulse waveforms with small relative errors, indicating the effectiveness of the model in reproducing the actual pulse pressures inside the vessel.
No takes yet. Share an insight, caveat, or question.
Jo et al. (2021) studied Pulse pressure measurement. Transfer function model for reconstructing radial pulse pressure waveforms vs. Internal measurement (built-in pressure sensor) and external on-skin measurement was evaluated on Accuracy of reconstructed internal pulse pressure waveforms. A transfer function model successfully reconstructed internal radial pulse pressure waveforms from external robotic tonometry measurements with small relative errors compared to actual internal pressures.