Key result
IVUS identifies stent compression as the primary cause of restenosis across all evaluated balloon-expandable stents.
Observational (n=11)
Answers are grounded in this paper and cite their sources.
IVUS reveals that stent compression is the principal cause of restenosis in balloon-expandable stents placed in compressible vascular sites like the SFA and hemodialysis conduits, a finding often missed by angiography.
OBJECTIVES: We sought to determine the basis for restenosis within superficial femoral arteries (SFAs) and hemodialysis conduits treated with balloon-expandable stents. BACKGROUND: Use of stents within coronary and peripheral vessels continues to increase exponentially. The mechanism of restenosis within stents placed at various vascular sites is not well understood. In particular, the implications of deploying a balloon-expandable stent in a compressible site are not well understood. METHODS: After the serendipitous detection of stent deformation during intravascular ultrasound (IVUS) examination of a restenosed dialysis fistula, we evaluated a consecutive series of patients with stents placed in compressible vascular sites, including the SFA (six patients) and hemodialysis fistulae (five patients). Clinical, angiographic and IVUS examinations were performed to evaluate mechanisms of restenosis. RESULTS: Stent compression was identified as the principal cause of restenosis in all dialysis conduits and SFAs. Stent deformity was not reliably identified by angiography; however, IVUS identified compression of two forms: eccentric deformation, implicating two-point compressive force, and complete circumferential encroachment of stent struts around the catheter, suggesting multidirectional compressive force. Despite redilation, secondary restenosis resulting from recurrent compression recurred in most sites. CONCLUSIONS: Restenosis within balloon-expandable endovascular stents may occur as a result of stent compression, a phenomenon readily detected by IVUS, but often not by angiography. These findings have significant implications for the use of balloon-expandable stents within vascular sites subject to extrinsic compression, such as hemodialysis conduits, the adductor canal segment of the SFA and carotid arteries.
No takes yet. Share an insight, caveat, or question.
Rosenfield et al. (1997) conducted an observational in Restenosis within superficial femoral arteries and hemodialysis conduits (n=11). Balloon-expandable stents in compressible vascular sites was evaluated on Mechanism of restenosis. Stent compression was identified by intravascular ultrasound as the principal cause of restenosis in all 11 dialysis conduits and superficial femoral arteries treated with balloon-expandable stents.
Synapse has enriched 10 related papers. Consider them for comparative context: