Los puntos clave no están disponibles para este artículo en este momento.
Five hundred and eighty-nine consecutive transvenous pacemakers implanted between January 1, 1971 and January 1, 1985 were reviewed. A complication incidence of 17% early in the study period (1971-1977) was caused by electrode displacement and perforation. As of 1975, smaller-sized electrodes with endocardial fixation were used and were associated with a significant decrease in this complication incidence to 2% by 1985. All other complications occurring during the entire study period (1971-1985) occurred at a frequency of 3% (20/589). Infection occurred in 3 of 589 cases; there were no extrusions or erosions. No mortality was attributed to pacemaker insertion. The "twiddler" syndrome seems to be increasing in frequency. Transvenous permanent pacing can be accomplished today with a low complication rate of 5% or less. The majority of these complications are minor and can be corrected easily. Our studies suggest that the reduced complication incidence is mainly related to improved technology.
Philip E. Hill (Fri,) studied this question.