BACKGROUND: The type of blood collection tube affects specimen quality and laboratory results. Because plasma specimens have a shorter processing time compared with serum specimens, emergency biochemistry tests use plasma. However, serum specimens remain stable after centrifugation and show more accurate results than plasma. Therefore, a quick-clotting serum separator tube is expected to be useful for shorter turnaround times and accurate results. We evaluated a new quick-clotting serum separator tube VQ-Tube™ (AB Medical, Korea) for clinical chemistry and thyroid hormone assays. METHODS: One hundred volunteers from four university hospitals were recruited, and peripheral blood samples were collected in quick-clotting serum separator tube VQ-Tubes™ and the commonly used serum separator tube V-Tubes™. The obtained specimens were used for 16 clinical chemistry assays and three thyroid hormone assays. RESULTS: -test revealed significant differences between the results of six assays, but each set of results showed a good correlation. Samples were visually inspected for serum clarity and gel barrier integrity, and incomplete clotting reactions and haemolysed serum were not observed. CONCLUSIONS: The new quick-clotting VQ-Tube™ demonstrated reliable test results compared with the commonly used serum separator tube V-Tube™. This quick-clotting tube will provide fast test results with adequately separated serum specimens, especially for patients who need fast tests.
Jo et al. (Tue,) studied this question.