April 6, 2010Classical and Quantum Gravity19 citationsOpen Access

Accurate calibration of test mass displacement in the LIGO interferometers

EGE. GoetzRSR. L. SavageJGJ. A. Garofoli

Key Points

Key points are not available for this paper at this time.

Abstract

We describe three fundamentally different methods we have applied to calibrate the test mass displacement actuators to search for systematic errors in the calibration of the LIGO gravitational-wave detectors. The actuation frequencies tested range from 90 Hz to 1 kHz and the actuation amplitudes range from 10-6 m to 10-18 m. For each of the four test mass actuators measured, the weighted mean coefficient over all frequencies for each technique deviates from the average actuation coefficient for all three techniques by less than 4%. This result indicates that systematic errors in the calibration of the responses of the LIGO detectors to differential length variations are within the stated uncertainties. © 2010 IOP Publishing Ltd.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Goetz et al. (2010) studied this question.

synapsesocial.com/papers/6a0864bf7de338f10b10a5c4https://doi.org/10.1088/0264-9381/27/8/084024