Key result
Tricuspid TEER induces significant inward-pulling annular forces, maximizing with an AS-PS two-clip intervention.
Why the study?
Tricuspid transcatheter edge-to-edge repair can induce an acute annuloplasty effect, but the mechanisms driving the reduction in annular size remain unclear.
Does the choice of leaflet pairs and number of clips in tricuspid TEER affect the magnitude of induced annular forces in an in vitro porcine heart model?
Does the choice of leaflet pairs and number of clips in tricuspid TEER affect the magnitude of induced annular forces in an in vitro porcine heart model?
In an in vitro porcine model, tricuspid TEER induces significant inward-pulling annular forces that vary based on the specific leaflet pairs clipped and the number of clips used.
Porcine TEER forces vary by clip strategy; leaves open whether these affect clinical outcomes or durability.
Background Tricuspid transcatheter edge-to-edge repair (TEER) can induce an acute annuloplasty effect. While this has a therapeutic benefit, the mechanisms driving the reduction in annular size remain unclear. Objectives We quantify the annular force induced by TEER in vitro in whole porcine heart preparations. We explore the impact of clipping different leaflet pairs on the TEER-induced annular forces. Methods We performed 49 interventions in 13 porcine hearts using a MitraClip XT. The clip was implanted between either the anterior-septal (AS), anterior-posterior (AP), or posterior-septal (SP) leaflet pairs. We also considered two-clip interventions between the combination of the AS-AP, AS-PS, or AP-PS leaflet pairs. For each intervention, we measured the right ventricular pressure, transvalvular flow rate, and force at eight locations around the annulus. Results TEER induced significant inward-pulling forces on the annulus. The maximum force was induced following an AS-PS two-clip intervention. A single AS clip induced the largest force among the one-clip interventions. Furthermore, the AP and AS-AP interventions induced the smallest annular forces. Conclusions The magnitude of the TEER-induced force depends on the intervention and number of clips implanted.
No takes yet. Share an insight, caveat, or question.
Haese et al. (2026) studied Tricuspid transcatheter edge-to-edge repair (n=13). Tricuspid transcatheter edge-to-edge repair (TEER) using MitraClip XT vs. Different clipping strategies (AS, AP, SP, AS-AP, AS-PS, AP-PS) was evaluated on Annular force, right ventricular pressure, and transvalvular flow rate. Tricuspid transcatheter edge-to-edge repair induced significant inward-pulling forces on the annulus, with the maximum force observed following an AS-PS two-clip intervention.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: