This is the authors' abstract. We don't add key points for this paper.
We demonstrate an approach to integrate multiple transducers on the tip of an optical fiber, leading to multiplexed and multiparameter sensing. The sensor consists of four photonic crystals that are placed on the tip of a cleaved 125 <tex-math notation="LaTeX">μ</tex-math>m diameter four-core fiber. Each photonic crystal has a different design which ensures that the sensitivities to temperature and refractive index of the solutions are different, enabling the simultaneous measurement of temperature and concentration. The sensor was calibrated in solutions of deionized water and ethylene glycol in a temperature range of <tex-math notation="LaTeX">25\,^∘</tex-math>C to <tex-math notation="LaTeX">65\,^∘</tex-math>C by fitting using a two-dimensional second order polynomial. Non-linear least square estimation was applied to determine concentration and temperature based on test data. The estimation has a high coefficient of determination where both parameters have a <tex-math notation="LaTeX">R²>0.998</tex-math>. The root mean square errors were found to be <tex-math notation="LaTeX">RMSET=0.45\,^∘</tex-math>C and <tex-math notation="LaTeX">RMSEc=0.20</tex-math>% for the temperature and concentration estimation respectively. The demonstrated sensor represents a scalable platform for multiplexed sensing on the tip of an optical fiber, with unprecedented spatial density. It could open the way to referenced biosensing to multiple analytes.