Key result
ACT measurement platforms differ by ~55 seconds, driving inconsistent real-world heparin dosing strategies.
Why the study?
ACT is used to guide intraprocedural heparin during left atrial ablation assuming device-independent targets, but whether measured ACT values differ across platforms and affect clinical practice was unknown.
Do measured ACT values and real-world anticoagulation strategies differ across measurement platforms during left atrial ablation?
Systematic Review (n=164)
Yes
Do measured ACT values and real-world anticoagulation strategies differ across measurement platforms during left atrial ablation?
Effect estimate: Mean difference 50-60 seconds
ACT values and real-world heparin dosing strategies vary significantly depending on the measurement platform used during left atrial ablation, highlighting the need for platform-specific interpretation.
Platform-specific ACT targets are required during left atrial ablation; challenges interchangeability assumptions and leaves optimal standardization open.
BACKGROUND: Activated clotting time (ACT) is used to guide intraprocedural heparin anticoagulation during left atrial ablation, with guideline-recommended target values generally assumed to be device-independent. OBJECTIVE: This study aimed to assess whether measured ACT values differ across measurement platforms and whether contemporary anticoagulation strategies vary accordingly in clinical practice. METHODS: We performed a systematic review of studies directly comparing ACT values across different measurement platforms during left atrial ablation. In parallel, we conducted an international physician survey assessing ACT platforms, heparin dosing strategies, and targeted ACT ranges. RESULTS: The systematic review identified 4 electrophysiology-specific studies comparing ACT platforms. Across all studies, simultaneously measured ACT values differed substantially among platforms, with mean paired differences of 50-60 seconds, corresponding to a 15%-25% discrepancy at ACT targets of 300-350 seconds. These interplatform differences persisted despite high correlation among devices. The survey was completed by 164 electrophysiologists (82% from Europe and 13% from North America). The most used platforms were Hemochron (42%), Medtronic ACT Plus (27%), and Abbott i-STAT (16%). Target ACT ranges and heparin dosing strategies varied by platform: higher ACT targets and higher weight-based heparin doses were most frequently reported with Hemochron; lower targets and doses were more common with i-STAT; and users of Medtronic showed an intermediate pattern. Nearly one-third of respondents reported modifying heparin dosing after switching ACT platforms. CONCLUSION: Available cross-validation data indicate substantial differences among ACT platforms. Real-world heparin administration strategies and ACT targets vary markedly by platform. These findings challenge the assumption of universal ACT targets and highlight the need for platform-specific interpretation.
No takes yet. Share an insight, caveat, or question.
Kovacs et al. (2026) conducted a systematic review in left atrial ablation (n=164). Activated clotting time (ACT) measurement platforms vs. Different ACT platforms was evaluated on ACT value differences across measurement platforms (Mean difference 50-60 seconds). Simultaneously measured ACT values differed substantially among platforms with mean paired differences of 50-60 seconds, leading to varied real-world heparin administration strategies.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: