PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
January 2, 1990Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE1 citations

Vacuum BRDF Measurements Of Cryogenic Optical Surfaces

View Full Paper
THTimothy L. HowardPGPatricia M. GeorgeSFSteve Flammang

Key Points

Key points are not available for this paper at this time.

Abstract

Spaceborne optical systems designed for long-duration missions are susceptible to degradation from contaminants in the spacecraft environment. In order to assess contamination build-up rates and corresponding scatter changes, we have designed a BRDF measurement system for use in a high-vacuum (10-8 Torr), low-temperature (20K) environment. The system measures cryocooled optics at angles from near-specular (< 1°) to approximately 30° off-axis, at wavelengths of 0.633 and 10.6 microns. We describe the design characteristics of the instrument, and discuss preliminary experiments which demonstrate contamination build-up correlated with changes in the sample BRDF profiles. The current status of the instrument is reviewed and future plans and applications summarized.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Howard et al. (1990) studied this question.

synapsesocial.com/papers/6a091e97b7dd28a06e15f89bhttps://doi.org/10.1117/12.962863
Ask AI
Helpful
Bookmark
Share
View Full Paper