Why the study?
Does a water-filled sphygmomanometer provide systolic blood pressure measurements that correlate with conventional cuffs while allowing ultrasonic imaging in elderly subjects?
Does a water-filled sphygmomanometer provide systolic blood pressure measurements that correlate with conventional cuffs while allowing ultrasonic imaging in elderly subjects?
A novel water-filled sphygmomanometer allows for simultaneous ultrasonic imaging of the brachial artery and provides systolic blood pressure measurements that correlate well with conventional cuffs.
Answers are grounded in this paper and cite their sources.
May enable simultaneous brachial imaging during systolic BP assessment; leaves open larger validation before clinical adoption.
A technique is described which enables ultrasonic imaging of the brachial artery whilst pressure is applied via a pressure cuff. This involves a new instrument--a sphygmomanometer, which uses water as opposed to air as the pressure medium, in order to permit ultrasonic imaging through the cuff. The technique was found to be acceptable in the clinical setting, and gave a measurement of the systolic blood pressure which correlated with the conventional cuff measurement in eleven elderly subjects (r = 0.89, p < 0.001). The technique should have an important role to play in studying the origin of differences which occur between direct and indirect blood pressure measurements.
No takes yet. Share an insight, caveat, or question.
Sheaham et al. (1993) studied this question.