OBJECTIVE: The purpose of this study was to evaluate the reproducibility (interreviewer agreement) and repeatability (intrareviewer agreement) of ROI sampling strategies to measure chemical shift-encoded (CSE) MRI-based liver proton density fat fraction (PDFF) and R2* (1 / T2*). A secondary purpose was to standardize ROI-based liver PDFF and R2* measurements by providing a compromise between measurement reproducibility and repeatability and time burden for image analysts. MATERIALS AND METHODS: CSE data from two cohorts were retrospectively analyzed. Cohort A included 53 patients referred for abdominal MRI and healthy subjects recruited for a comparison study of CT and MRI. Cohort B included 37 patients with suspected liver iron overload. Three reviewers measured liver PDFF and R2* using previously reported ROI sampling strategies. Inter- and intrareviewer agreement of liver PDFF and R2* were evaluated using Bland-Altman analysis. RESULTS: for cohort B. This approach was the most time-burdensome, requiring a mean ± SD of 149.7 ± 8.6 s per dataset. CONCLUSION: For improved reproducibility and repeatability of liver PDFF and R2* measurements, clinicians and researchers should sample as much area of the liver as possible using multiple large ROIs.