The electrochemical immobilization of salicylate hydroxylase (SH) within poly(amphiphilic pyrrole ammonium) films electrogenerated on carbon electrode is described. The amperometric determination of salicylate is carried out at 0.4 V (vs. Ag/AgCl) via the oxidation of the enzymically generated catechol. The detection limit and the sensitivity of the polypyrrole-SH electrode were 3×10−9 M and 1.7 A M−1 cm−2, respectively. Moreover, the co-entrapment of SH and polyphenol oxidase (PPO) in such polypyrrole films provides a bienzyme electrode also for the determination of salicylate. Contrary to the monoenzyme electrode, the functioning principle of the bienzyme electrode is based on the reduction of o-quinone enzymically generated from catechol. Consequently, the amperometric detection of salicylate was performed at −0.2 V without interferences from easily oxidizable compounds. The calibration range for salicylate measurement and the biosensor sensitivity were 5×10−8–3×10−6 M and 0.25 A M−1 cm−2, respectively.
No takes yet. Share an insight, caveat, or question.
Cosnier et al. (2001) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: