Key result
Automatically planned cardiovascular magnetic resonance acquisitions yielded accurate quantification of global left ventricular function, with interstudy factors accounting for 70-85% of total variation.
Why the study?
Does automated CMR scan planning with positional and angular variation affect the quantification of left ventricular dimensions and function compared to manual planning in subjects undergoing CMR?
Observational (n=19)
Does automated CMR scan planning with positional and angular variation affect the quantification of left ventricular dimensions and function compared to manual planning in subjects undergoing CMR?
Inaccurate positioning and tilting of short-axis images in automatically planned CMR do not significantly affect the quantitative measurements of LV function.
Automated CMR planning supports reliable LV function quantification despite variations; hypothesis-generating for routine use pending prospective validation.
PURPOSE: To theoretically and experimentally investigate the influence of the automated cardiovascular magnetic resonance (CMR) scan planning pitfalls, namely inaccurate positioning and tilting of short-axis (SA) imaging planes, on quantification of the left ventricular (LV) dimensions and function. MATERIALS AND METHODS: Eleven healthy subjects and eight patients underwent CMR. Manually and automatically planned SA sets were acquired. To obtain the quantitative measurements of LV function, one observer performed image analysis twice. The agreement between planning methods, as well as the decomposition of the total variation into interstudy and intraobserver components was measured. RESULTS: The decomposition of the total variation showed that the interstudy factor accounts for 70-85% of the total variation, while the rest is due to the intraobserver factor. Moreover, the relative contribution of the interstudy factor remains independent from errors in slice positioning and small angular deviation of SA stacks from the optimal orientation. Good agreement between the theoretical and measured variability factors was observed. CONCLUSION: Global LV function derived from the automatically planned CMR acquisitions yield accurate quantification of the human cardiovascular system. Inaccurate positioning and tilting of SA images does not affect the quantitative measurements of LV function. The computer-aided system for automated CMR has proven clinical applicability.
No takes yet. Share an insight, caveat, or question.
Danilouchkine et al. (2005) conducted an observational in Healthy subjects and patients undergoing CMR (n=19). Automatically planned short-axis (SA) imaging planes vs. Manually planned SA sets was evaluated on Agreement between planning methods and decomposition of total variation into interstudy and intraobserver components for LV function quantification. Automatically planned cardiovascular magnetic resonance acquisitions yielded accurate quantification of global left ventricular function, with interstudy factors accounting for 70-85% of total variation.