Key result
Brachial oscillometry measures MAP ~6 mmHg higher than radial tonometry by utilizing fixed form factors.
Why the study?
Mean arterial pressure is often estimated using a fixed arterial form factor, but waveform shapes vary and true measured form factor should vary, potentially affecting mean arterial pressure accuracy.
How do radial tonography and brachial oscillometry compare in estimating arterial form factor and mean arterial pressure?
Observational (n=533)
No
How do radial tonography and brachial oscillometry compare in estimating arterial form factor and mean arterial pressure?
Absolute Event Rate: 96.3% vs 102.6%
p-value: p=0.000
Arterial form factor varies significantly based on measurement site and device, with oscillometric devices using a fixed form factor that may overestimate mean arterial pressure compared to tonometry.
Device MAP discrepancies warrant caution in interchangeable use; leaves open standardization of form factor for central pressure estimation.
Mean arterial pressure (MAP) is often estimated from cuff systolic (S) and diastolic (D) blood pressure (BP) using a fixed arterial form factor (FF, usually 0.33). If MAP is measured accurately, a true FF can be calculated: FF = [MAP–DBP] / [SBP–DBP]. Because waveform shapes vary, true measured FF should also vary and MAP accuracy may be affected. We studied factors affecting FF using radial tonography (SphygmoCor, n = 376) or brachial oscillometry (Mobil-O-Graph, n = 157) and to compare devices, 101 pairs were matched precisely for SBP and DBP. SphygmoCor brachioradial FF correlated strongly with central FF (r 2 = 0.75), central augmentation index (cAI, r 2 = 0.39), and inversely with pulse pressure amplification (PPA) ratio (r 2 = 0.44) [all p < 0.000]; brachioradial FF was lower than central (c) FF (0.34 vs 0.44, 95% CI’s [0.23,0.46] and [0.34,0.54], p < 0.000); and on forward stepwise regression, brachioradial FF correlated with PPA ratio, age, heart rate, and cAI (multiple-r 2 0.63, p < 0.000). With Mobil-O-Graph: brachial FF was fixed, lower than the corresponding cFF [mean(SD)] 0.46(0.00098) vs 0.57(0.048), p < 0.000], and uncorrelated with other characteristics; MAP and cSBP were higher than SphygmoCor by 6.3 and 2.2 mmHg (p < 0.005) but also had significant systematic negative biases. We conclude that with radial tonometry (SphygmoCor), FF varies with pulse wave morphology within and between individuals and by measurement site, age, and heart rate. With oscillometry (Mobil-O-Graph), brachial FF was unrelated to other clinical variables, fixed and high; MAP and cSBP were higher than tonometry, with systematic negative biases.
No takes yet. Share an insight, caveat, or question.
Izzo et al. (2022) conducted an observational in Normotension and Hypertension (n=533). Radial tonometry (SphygmoCor) vs. Brachial oscillometry (Mobil-O-Graph) was evaluated on Mean arterial pressure (MAP) in matched pairs (p=0.000). Brachial oscillometry (Mobil-O-Graph) produced a mean arterial pressure 6.3 mmHg higher than radial tonometry (SphygmoCor) in matched pairs, utilizing a fixed rather than variable form factor.