Key result
A modified quadrupole gradient set achieved gradient field strengths up to 15.2 G/cm with field distortions less than 2% over the central 8 cm, suitable for high-resolution MRI tagging.
A custom quadrupole gradient set enables high-resolution MRI tagging of isolated canine hearts in a standard clinical 1.5 T scanner.
Advances preclinical cardiac tagging; leaves open translation to in vivo human imaging.
A special purpose gradient set with a Cos(2 theta) current distribution has been constructed for high resolution MR imaging of small samples in a clinical Signa 1.5 T scanner using the existing gradient amplifiers. The X, Y, and Z gradient coils can attain gradient field strengths of 11.3, 4.7, and 15.2 G/cm at 100 amps current, respectively, with a slew rate of 20 G/cm/ms and usable ramp time of 150 microseconds. Field distortions are less than 2% over the central 8 cm of the bore, suitable for high resolution tagging of isolated canine hearts.
No takes yet. Share an insight, caveat, or question.
O’Dell et al. (1994) studied this question. Modified quadrupole gradient set was evaluated on Gradient field strengths and field distortions. A modified quadrupole gradient set achieved gradient field strengths up to 15.2 G/cm with field distortions less than 2% over the central 8 cm, suitable for high-resolution MRI tagging.