Cadmium zinc telluride (CZT) detectors are suitable for various applications due to the good energy resolution and the simple pixilation to achieve high spatial resolution. Our group is developing a two-panel head-and-neck dedicated positron emission tomography system based on CZT detectors. Each panel will consist of 150 CZT crystals (4 <tex-math notation="LaTeX">× </tex-math> 4 <tex-math notation="LaTeX">× </tex-math> 0.5 cm 3 ) covering an area of 20 <tex-math notation="LaTeX">× </tex-math> 15 cm 2 in an edge-on configuration to achieve high detector efficiency at 511 keV. In this work, we present the design and development of a full data acquisition (DAQ) chain that enables a low noise and compact readout for each panel. The initial results of the readout circuit were quantified using a 1-kHz square wave test pulse. The pulse amplitude was chosen to generate approximately the same amount of charges as a 511-keV photon would provide in CZT. The best-case full width at half maximum (FWHM) electronic noise at 511 keV was measured to be 0.69% ± 0.16% (3.52 ± 0.81 in keV units after conversion). The FWHM electronic noise at 511 keV for a complete DAQ chain was 4.33% ± 0.30% (22.13 ± 1.53 in keV units).