Portico valve recipients experienced a more profound drop in normalized platelet counts at 3 days post-implantation compared to Evolut recipients (60.2 vs 71.8), which was mitigated by extra rinsing of the storage solution.
Observational (n=238)
No
Does the choice of transcatheter aortic valve device (Portico vs Evolut) affect periprocedural platelet damage and pro-inflammatory state in TAVI recipients?
The formaldehyde-based storage solution of the Portico valve may cause platelet injury and thrombocytopenia, which can be mitigated by additional flushing prior to implantation.
Absolute Event Rate: 60.2% vs 71.8%
Thrombocytopenia after TAVI is common and clinically detrimental. Retrospectively, we observed Portico recipients had a more profound platelet drop than Evolut recipients. We thus investigated periprocedural platelet damage and/orpro-inflammatory state in 64 TAVI recipients at baseline and after implantation. Platelet damage was assessed by annexin V staining and monocyte-phagocytic phenotype was assessed according to CD14/CD36 expression. Serum cytokines were measured in 20 patients. The formaldehyde-based storage solution altered platelets. When, before being loaded onto the delivery system, Portico underwent one additional flushing to those recommended, the receiving patients showed thrombocytopenia, platelet damage, and CD36-monocyte count were mitigated. A general increase in IL-6 was recorded in overall TAVI recipients, but a high serum level of IL-8, a potent thrombocytopenia inducer, was measured in Portico recipients only, including those with extra-rinsed valve. Our study suggests a platelet-injury effect by storage-solution and generates the hypothesis of a role for the biomaterial in stimulating innate-immunity. Larger prospective studies are needed. Graphical Abstract.
Corcione et al. (Mon,) conducted a observational in Severe aortic valve stenosis (n=238). Portico transcatheter aortic valve vs. Evolut transcatheter aortic valve was evaluated on Normalized platelet count at 3 days post-implantation. Portico valve recipients experienced a more profound drop in normalized platelet counts at 3 days post-implantation compared to Evolut recipients (60.2 vs 71.8), which was mitigated by extra rinsing of the storage solution.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: