We present an optimal design of an X-ray target for 6 MeV medical linear accelerators using FLUKA simula-tions. The target is composed of high-atomic number tungsten and high-thermal conductivity copper, corre-sponding water-cooling system is showed too. Further-more, we analyse the temperature and thermal stress re-sponses of the target under transient thermal loads using Ansys Code. For 6 MeV electron beam with 100 uA cur-rent, the results show that the target can achieve 1014 cGy/min at 1meter in front of the target. Within 100 ms, the maximum temperature reaches 512 °C under pulsed heating source with 250 Hz frequency and 1' duty cycle and the number of cycles to failure is estimated as 5.8·10⁸.